Planmeca is one of the world's leading manufacturers of digital dental units, CAD/CAM solutions, 2D and 3D imaging devices and software solutions. User’s Manual 4 PLANMECA INTRA X-RAY UNIT - MAIN PARTS 4.1 General view of the X-ray Horizontal tube head stiffness adjustment screw Adjust the stiffness of the tube head horizontal movement by turning the adjustment screw on the support axle manually or with a wrench tool (see next page). Planmeca Ec Proline Manual Planmeca Proline Ec Service Manual Planmeca Proline Xc Service Manual Planmeca Pn 2002 Cc Proline Ec Proline Pm 2002 62038 Proline Proline Pm 2002 Cc Pm 2002 Cc Proline Download: Ec Proline Pm 2002 Isuzu N Series Diesel Owners Manual / Drivers Manual Guide Book 2016 Ford Tractor 2600 3600 4100 4600 Operators Manual - Original Manual 2007 Manual Mini Cooper.
Planmeca 2002 cc proline error
Planmeca intra x-ray unit 3 warnings and precautions user’s manual 2 warnings and precautions note it is very important that the place where the unit is to be used and the position from which the user is to operate the unit are correctly shielded. Since radiation safety requirements vary from country to country and state to state it is the. Planmeca Proline Ec Service Manual Planmeca Proline Xc Service Manual Planmeca Ec Proline Manual Planmeca Pn 2002 Cc Proline Proline Pm 2002 Cc 62038 Proline Ec Proline Pm 2002 Pm 2002 Cc Proline Download: Ec Proline Pm 2002 Seloc Service Manuals Seloc Service Manual Yamaha All 4-stroke 2005-10 Chinese 125 Motorcycles Service And Repair Manual. Planmeca Intra User Manual User's manual Configuring Planmeca mRomexis server. Configure your Planmeca Romexis server to listen to Planmeca mRomexis (port 8093. User Guides, Installation Guides, Service Manuals, Brochures ADEC 532 / 533 / 542 Delivery System User Guide Planmeca Intra X-Ray User Manual. Welcome to Planmeca! Planmeca Oy is one of.
September 02, 2014 10:10
Hello, I've the Planmeca Proline 2002 cc Ortho and the machine reports me the Error 57 in the display. I've turn off and then turn on the Ortho, but the error continues.
Can anyone help me?
Thanks!
Can anyone help me?
Thanks!
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
remove the handswitch and turn unit off and on and see if it clears? could be a bad switch closed all the time
Log inor Registerto rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
Error 57 hand switch activated during switch on check this first
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
Log inor RegisterKeyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
to rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:02
Hi, you have problem in exposure button, praticaly is switched on. Open the exposure botton and also you can have problem in the extention cable. Put out the cable of extention exposure button and the error must deasapeare.
Send me a mail, and i send you technical manual.
Boian
Log inor RegisterSend me a mail, and i send you technical manual.
Boian
to rate and post a comment
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
re: re: re: Planmeca 2002 cc proline error
September 03, 2014 05:34
HE 2 Amalgam container is 100% full (only Metasys
or Dürr).
Replace or empty the container immediately.
The suction system cannot be used before the
container is emptied or replaced. See the
User’s manual for details.
send me an email address I will send you the service manual, if your keyboard does not respond and is used to initate the exposure it probably is bad
Log inor Registeror Dürr).
Replace or empty the container immediately.
The suction system cannot be used before the
container is emptied or replaced. See the
User’s manual for details.
send me an email address I will send you the service manual, if your keyboard does not respond and is used to initate the exposure it probably is bad
to rate and post a comment
re: re: re: re: Planmeca 2002 cc proline error
September 04, 2014 10:19
Amalgam in x-ray?!?
Log inor Registerto rate and post a comment
re: re: re: re: re: Planmeca 2002 cc proline error
September 05, 2014 09:38
that came from the service manual
re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 10:08
Hello Lenny, i need this service manual, can you send it also to me please [email protected]
Thanks so much
Thanks so much
re: re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 10:33
Planmeca Proline Xc Calibration
OK I will send
Log inor Registerto rate and post a comment
re: re: re: re: re: re: Planmeca 2002 cc proline error
March 10, 2017 04:46
Thanks a lot
One moment please...
DESCRIPTION
Proline XC X-rays are designed for full-view patient positioning...Proline XC X-rays are designed for full-view patient positioning using a three laser positioning light system and a Graphic User Interface with easy-to-follow, color-coded menu options. The XC's focal trough adjusts to the patient's jaw shape and size providing excellent image quality. In addition, the Proline XC can be upgraded with future software features by simply replacing a software chip! With over 10+ programs from which to choose, the XC is definitely an economical solution for excellent dental imaging.
FORUMSView All (8)
Ask a New Question0Replies10 months ago | 10 months agotroubleshooting message on computer after taking pano I took a pano today and the pano worked but did not come up on the computer and the message says no license on fileReply |
-Del Mar View Dental Care 2 years ago | 2 years agoit wont rotate properly, off center at start up. It wont finish it wont rotate properly, off center at start up. It wont finishReply |
0Replies3 years ago | 3 years agoPlanmeca ProMax We have a Promax that the images produces a white line at mid-line during exposure. We are exporing possible solutions to the white out at mid-line. Reply |
DOCUMENTS / MANUALSView All
VIDEOS
FEATURES
- Open and easy patient access
- Comfortable and stable patient supports
- Side entry and open view for practical and precise patient positioning
- Triple laser beam system for accurate alignment of reference anatomical landmarks
- TheGraphicUserInterface(GUI)for intuitive selection of exposure program and parameters
- With state-of-the-art direct digital imaging the image is available for diagnosis immediately after exposure
Additional Specifications
Generator: Constant potential, microprocessor controlled, operating frequency 80 kHz
X-ray tube: D-052SB
Focal spot size: 0.5 x 0.5 mm
Total filtration: 2.5 mm Al
Anode voltage: 60 - 80 kV
Anode current: 4 - 12 mA DC
Exposure time:
Pan-2.5 - 18 s
Ceph- 0.2 - 5 s
X-ray tube: D-052SB
Focal spot size: 0.5 x 0.5 mm
Total filtration: 2.5 mm Al
Anode voltage: 60 - 80 kV
Anode current: 4 - 12 mA DC
Exposure time:
Pan-2.5 - 18 s
Ceph- 0.2 - 5 s